Define the measurand
Agree the material, wet or dry basis, measurement volume and decision to be supported.
Quality and validation
A credible moisture project links every displayed value to a defined reference method, representative samples and documented acceptance conditions.

Quality and validation
Skipping one link can make a stable-looking value unsuitable for control, quality release or settlement.
Agree the material, wet or dry basis, measurement volume and decision to be supported.
Use the applicable laboratory, oven or recognised industry method and record its conditions.
Match online time and location while protecting samples from moisture loss or gain.
Cover real moisture, density, temperature, formulation, particle size and seasonal variation.
Relate sensor response to reference results with a controlled model and version record.
Test data not used to build the model and compare results against pre-agreed criteria.
Monitor bias, process changes, maintenance and the triggers for revalidation or recalibration.
Quality and validation
Each workstream answers a different risk. Together they define whether the result remains valid in the real process.
Confirm method scope, sample preparation, drying or laboratory conditions, calculation basis and current revision.
Cover space, time, depth, batch, supplier and process state; pair each sample with the correct online interval.
Include the full intended range and influential variables; separate model-building data from independent validation.
Review bias, residuals, repeatability and errors by material group and operating range—not one headline number.
Document configuration, I/O, installation, communications, paired sampling, alarm logic and agreed acceptance tests.
Use routine comparison, trend review, change records and clear triggers for adjustment or full recalibration.
Quality and validation
The package should allow another engineer or quality manager to understand what was tested, under which conditions and where the limits are.
Published performance must remain tied to the tested material and installation conditions.